4-2 Basic Settings and Advanced Settings

4-10

4

Setting V

a

ri

o

us
F

u

nct
ion
s

IG-E

Outer diameter/Width measurement mode

The internal measurement value (R.V.) is determined by
measuring the outer diameter or width of the object.

„ Details

The distance from the top most light to dark transition as seen by the receiver to the bottom
most dark to light transition as seen by the receiver is determined as an internal
measurement value (R.V.).

Reference

• If there is no object present , the internal measurement value (R.V.) becomes

[

].

"2 Display Screen and Output" (page A-9)

• If the object is in the following status, the internal measurement value (R.V.)

becomes [

].

Measurement
range

R.V.

Top

Bottom

R.V.

Top

Bottom