Setting method 1 (one-point calibration), Setting method 1 (one-point calibration) -53, F'(.6 – KEYENCE IG Series User Manual

Page 111: 5 calibration function, Reference, Ig-e, Setting v a ri o us f u nct ion s, Main screen, When w

Advertising
background image

4-5 Calibration Function

4-53

IG-E

4

Setting V

a

ri

o

us
F

u

nct
ion
s

Setting Method 1 (One-point Calibration)

Main screen

1

Press [MODE] button and

X button for approx. 2

seconds on the main screen at the same time.

[

.END] is displayed on the main display (upper level).

[MODE] +

X button

2 seconds or more
at the same time

2

Press

S / T button to set [WUT] and press X button.

X button

3

Press the [SET] button without the object. (SET 1

Confirmation operation)

After [

UGV] blinks on the main display (upper level) several

times, [

UGV] is displayed.

[SET] button

4

Press

S / T button to set "R.V. you wish to display"

for [

UGV].

[

UGV] and the current internal measurement value (R.V.) are

displayed alternately on the main display (upper level).

5

Press the [SET] button while the "R.V. before

correction" is displayed on the main display (upper

level). (SET 2 Confirmation operation)

After [

UGV] blinks on the main display (upper level) several

times, the setting is complete and the main screen is restored.

[SET] button

Reference

• When

W

button is pressed, the display returns to the previous screen.

• When the [MODE] button is pressed or both the [MODE] button and

X

button are continuously pressed at the same time, the setting is canceled
and the main screen is restored.

• When

X button is pressed in step 4, the setting is canceled and the main

screen is restored.



LASER

BANK

0

1

2

3

HI

GO

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT



HOLD

CALC

CHECK

F'(.6

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

%.D

GO

HOLD

CALC

CHECK

Calibration setting

Setting value

Description

Default value

FGHNV

Standard

c

WUT

User setting



LASER

BANK

0

1

2

3

HI

GO

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

5'6

HOLD

CALC

CHECK



LASER

BANK

0

1

2

3

HI

GO

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

5'6

HOLD

CALC

CHECK

Set the value you wish to display

Item

Setting range

Default value

UGV

-99.999 to 99.999

5.000

Main screen

Advertising