Self-timing Hold

The LS-7600 Series in self-timing hold operation will detect targets automatically with
no external timing input. Then the measured value is displayed and output.

Judgment

Measured value

Measured value

Measured value

Measured value

Judgment

t

Sampling

period

Sampling

period

ST

time

ST

time

No internal

measured value

No internal

measured value

Internal measured value

Internal measured value

Waiting for judgment

(---------- displayed)

Measured value

ON

OFF

RESET input

ON

OFF

Comparator output

ON

OFF

ON

OFF

BCD output

RS-232C output

Judgment awaiting output

(Comparator waiting output)

Strobe output

Warkpiece removed

Workpiece insertion

Workpiece insertion

Internal
measured value

• When a workpiece is within the sensing distance, the emergence of the internal measured value will be

used as a trigger. Then data will be sampled for the averaging number after the preset self-timing (ST)
period is over. The determined internal measured value will be held. Then the value will be output.

• In the following cases, the LCD monitor will display the result waiting status “---.-----“ until the first measured

value is determined.

The LS-7600 Series is turned on, setting changes are made, program number changes are made,
RESET input is ON, or auto zero is input.

• Comparator output turns ON according to the measured value.
• In the above chart, the ON status of each output means that NPN open collector output is ON, provided

that it is normally open type. If it is normally closed type, the signal will be inverted.

• The ON status of each input means that the input is short-circuited to the COM terminal.
• RS-232C output is ON in synchronization with timing input if the D-SEND item as an environment setting is

set to S1 or S2 only. Refer to Data Transmission Mode on page 6-9 for environment settings in detail.

• Timing input is ignored.
• If the internal measured value during a sampling period becomes “---.-----,” the LCD monitor will display the

result waiting status “---.-----.”

• RESET input can be under RS-232C control. Refer to page 8-11 for details.
• Strobe output time can be changed. Refer to Strobe Output Time on page 5-

47 for details.

• The LCD monitor will display “HOLD” until the next workpiece is located in

the measuring area after the present measured value is determined.