7-9

Chapter 7 I/O Terminals

7

BCD Output

Name

Description

Reference page

7-6

7-6

7-6

7-6

7-6

BCD POLE output

BCD polarity output.

OFF: Positive
ON: Negative

BCD DIGIT output

BCD numeric output for each digit.
Negative logic (i.e., Turns ON (set to 1) when NPN
open connector output is ON).

Comparator waiting status (with "–" displayed for
each digit): All digits set to B (1011)
Overrange (with "F" displayed for each digit): All
digits turn ON (1111).

BCD INDICATE
OUT1/2 output

Present OUT number turns ON.

OFF : OUT1
ON : OUT2

BCD SELECT
OUT1/2 input

Select and input the OUT number to be output.

OFF (Open) : OUT1
ON (Short-circuited to COM2 terminal) : OUT2

STROBE output

Output in synchronization with determined BCD
output.

Statistical Processing I/O

Name

Description

Reference page

STATS input

STATS RUN output

7-3, 7-5, 7-6

7-5

Statistical processing.

Output while in statistical process.

Electrical Specifications

The following section provides information on the I/O electrical specifications and circuitry
of the LS-7600 Series.

Non-voltage Input

Input

+5V

+5V

470

Ω

2.2k

Ω

4.7k

Ω

1k

Ω

COM1

/COM2

Inter

nal circuit

10nF

ON voltage

OFF current

Short-circuit current

(typical)

1 V max.

0.6 mA max.

2mA