5-26

Chapter 5 Function Settings

5

Two-point Calibration Settings

Make two-point calibration settings as explained below.

Setting Range

Two-point calibration requires two types of master workpieces that are different to each
other in size. If the master workpieces are target 1 (T1) and target 2 (T2) respectively,
make settings so that the following conditions will be satisfied.
T1-A < T2-A
Furthermore, set the calibration span between 0.5 and 2.0.

Note:

In the case of two-point calibration, the number of averaging measurements is
fixed at 4,096 times.

Press the [ENT] key when the cursor is at 2-TARGET.

1

Place the small master workpiece (T1-A) in the beam.

2

While the cursor is at T1-B, press the [ENT] key. Monitor the measured value
and input the T1-B value.

The display “*********” changes to “+00.00000.”
The cursor moves to the “+” position and to the value can be input.

(1) Press the [Up or Down] key and

select + or – as desired.

(2) Press the [Left or Right] key and

move the cursor between digits to
input the value.

AREA1

2-TRAGET

T1-A

0 .00000

T1-B

+

00 .00000

T2-A

✽✽✽✽✽✽✽✽✽

T2-B

✽✽✽✽✽✽✽✽✽ SET

5

0 0 3 0 0

AREA1

2-TRAGET

T1-A

0 .00000

T1-B

+0

5

.00000

T2-A

✽✽✽✽✽✽✽✽✽

T2-B

✽✽✽✽✽✽✽✽✽ SET

5

0 0 3 0 0

(3) Press the [Up or Down] key to

input the value.

(4) Repeat steps (2) and (3) until all

the digits are input.

(5) Press the [ENT] key.

When the input value is entered, the
measured value of the master
workpiece is set for the T1-A. Then
the cursor moves to T1-B.

3

Place the large master workpiece (T2-A) in the beam.

4

Press the [Down] key and select T2-B. Then input the value in the same man-
ner in step 2.

5

Press the [Down] key and move the cursor to SET.