Offset – KEYENCE LS-7600 User Manual

Page 105

Advertising
background image

5-43

Chapter 5 Function Settings

5

Offset

An offset value is a desired value that can be added to or subtracted from measured
values. If the offset value is set, it is possible to display the offset value immediately on the
screen by using the auto zero function.

Setting Range

The range is initially set to –99.99999 to 99.99999mm. This range varies with the
display unit or minimum display unit setting. Refer to Display Unit on page 5-57 and
Minimum Display Unit on page 5-58.
Input positive or negative values to increase or decrease the set value.

Setting Method

1

Press the [Up or Down] key and move the cursor to OFFSET.

2

Press the [ENT] key .

The cursor moves to +, where numeric input is possible.

3

Press the [Up or Down] key and
select + or –.

4

Press the [Left or Right] key and
move the cursor between digits to
input the value.

5

Press the [Up or Down] key and
select the value.

6

Repeat steps (4) and (5) until all the
digits are input.

7

Press the [ENT] key .

All the input items are entered and the
cursor returns to OFFSET.

OUT1

CALC

+1

A1

+0

A2

AVE

512

MEA-MODE

NORMAL

OFFSET

+

00.00000

OUT1

CALC

+1

A1

+0

A2

AVE

512

MEA-MODE

NORMAL

OFFSET +0

5

.00000

Master Calibration with Automatic Zero (Automatic Offset) Function

The automatic zero function sets measured values to the preset size values (i.e., offset
values) of master workpieces.
Refer to Measurement of Outer Diameter and Width (with Two Measuring Heads for
Larger Objects) on page 3-4 for an application example. For the details of the automatic
zero function, refer to Automatic Zero Function on page 4-10.

Advertising