5-11

Chapter 5 Function Settings

5

T-EDGE (Top Edge)/B-EDGE (Bottom Edge)

Select either of these settings for gap measurement or positioning of rollers.

When the LS-7600 Series is in T-EDGE measurement operation, it measures the width of
the bright portion between the top of the measuring head and the first shadow in the
measuring area.

When the LS-7600 Series is in B-EDGE measurement operation, it measures the width of
the bright portion between the bottom of the measuring area and the bottom of the last
object in the beam.

Target

Transmitter

Receiver

This space is measured in
T-EDGE mode or B-EDGE
mode.

Transmitter

Target

Receiver

This space is measured in
T-EDGE mode.

This space is measured in
B-EDGE mode.

Tips

Setting the reference edge within the measuring area enables measurements

in higher repeat accuracy.

SEG (Segment)

Select this setting for the measurement of the inner diameters of discs or the measure-
ment of pitches. The edge where the measurement of the object in the measuring area
starts is called “start edge” and the edge where the measurement terminates is called a
“end edge.” The space between these edges specified is measured. Segment measure-
ment is performed in E mode or P mode. If objects are measured with the virtual center
of the light axis set, use edge 0.

Setting Range
• Edge or pitch numbers can be designated between –127 and 127.
• The duplicated designation of an edge or pitch number will set the measured value to

0.

• If edges or pitches are designated in reverse order, negative measured values will be

obtained.