Temperature characteristics – KEYENCE LS-7600 User Manual

Page 199

Advertising
background image

9-5

Chapter 9 Specifications

9

LS-7070M

150

56

40

65

Receiver

Transmitter

Unit : mm

20

30

100

±

50

±

6

μ

m

±

3

μ

m

Temperature Characteristics

The following table shows typical examples of the temperature characteristics of the
LS-7600 Series.

LS-7010M

Temperature

Drift

0˚C (32˚F)

0.25

50˚C (122˚F)

-0.15

10˚C (50˚F)

0.10

20˚C (68˚F)

0.00

40˚C (104˚F)

-0.10

30˚C (86˚F)

-0.05

Unit:

μ

m

The above values are checked with a 1 round bar located in the center of the meas-
uring range at a reference temperature of 20˚C.

LS-7030M

Temperature

Drift

0˚C (32˚F)

–0.5

50˚C (122˚F)

0.2

10˚C (50˚F)

–0.1

20˚C (68˚F)

0.0

40˚C (104˚F)

0.1

30˚C (86˚F)

0.1

Unit:

μ

m

The above values are checked with a ø10 round bar located in the center of the
measuring range at a reference temperature of 20˚C.

LS-7070M

Temperature

Drift

0˚C (32˚F)

+1.8

50˚C (122˚F)

-3.0

10˚C (50˚F)

+0.8

20˚C (68˚F)

0.0

40˚C (104˚F)

-1.8

30˚C (86˚F)

-0.8

Unit:

μ

m

The above values are checked with a ø20 round bar located in the center of the
measuring range at a reference temperature of 20˚C.

Advertising