Calibration settings, Types of calibration settings – KEYENCE LS-7600 User Manual

Page 84

Advertising
background image

5-22

Chapter 5 Function Settings

5

Calibration Settings

The measured value of an target may have a delicate error due to the surface condition
and angle of the target. Calibration settings correct for this error. It is possible to make
calibration settings for each area independently.

Types of Calibration Settings

Select the type of calibration from the following ones in advance. Refer to Calibration
Type Selection on page 5-24.
Logic calibration:

Logic calibration requires two types of master workpieces that are
different to each other in size. Calibration is performed according
to pre-calibration and post-calibration values that are input by the
user. This type of calibration is used for the measurement of the
distance between two workpieces, for example.

Two-point calibration: Two-point calibration requires two types of master workpieces

that are different to each other in size.

One-point calibration: One-point calibration requires a single master workpiece. This

calibration is used for the size measurement of a single type of
workpiece.

Logic Calibration/Two-point Calibration

Prepare target 1 (T1) and target 2 (T2) as
two measuring points. The LS-7600 Series
performs calibration on these points based
on the T1-A and T2-A display values, which
are obtained before calibration, and the T2-
B and T2-B display values, which are
obtained after calibration. In this case, the
non-calibrated display values are converted
into the calibrated display values with the
span value (inclination) and shift value of
the workpiece determined by the LS-7600
Series internally.

T1

(1st point)

T2

(2nd point)

Workpiece
size

Display value

T2-B

(T2 after calibration)

T2-A

(T2 before calibration)

T1-B

(T1 after calibration)

T1-A

(T1 before calibration)

Ideal straight
line

Display value
after calibration

Display value
before calibration

T1

(1st point)

Workpiece
size

Display value

T1-B

(T1 after calibration)

T1-A

(T1 before calibration)

Ideal straight
line

Display value
after calibration

Display value
before calibration

One-point Calibration

Prepare target 1 (T1) as a single measuring
point. The LS-7600 Series performs calibra-
tion on this point based on the display value
before calibration and that after calibration.
In this case, the non-calibrated display
value is converted into the calibrated display
value with the span (inclination) and shift
value of the workpiece determined by the
LS-7600 Series internally. The span value is
always set to 1.