Calibration type selection, Logic calibration settings – KEYENCE LS-7600 User Manual

Page 86

Advertising
background image

5-24

Chapter 5 Function Settings

5

Calibration Type Selection

The following section provides information on how to set the calibration type for each area.
The procedure below is for AREA1 for a typical example.

1

With the AREA 1 box selected, press ENT. T1-A will be lit. Press the [Up] key
and move the cursor to underline the type of calibration. “THEORY”

The item is underlined and displayed.

2

Press the [ENT] key.

The cursor appears on “THEORY”.

3

Press the [Up or Down] key and
select the type of calibration for
AREA1.

Any one of the following items displayed
can be selected.

Logic calibration

THEORY

Meaning

LCD monitor

Two-point calibration

2-TARGET

One-point calibration

1-TARGET

Initialization of calibration

RESET

AREA1

THEORY

T1-A

✽✽✽✽✽✽✽✽✽

T1-B

✽✽✽✽✽✽✽✽✽

T2-A

✽✽✽✽✽✽✽✽✽

T2-B

✽✽✽✽✽✽✽✽✽ SET

4

Press the [ENT] key.

5

To enable the type of calibration that is selected, calibrate the LS-7600 Series
with the right method. For details, refer to the settings for each type of calibra-
tion.

Tips

Take steps 1 through 4 to select the type of calibration for AREA2 as well.

Refer to page 5-28 for the initialization of calibration.

Logic Calibration Settings

Make logic calibration settings as explained below.

Setting Range

Logic calibration does not require master workpieces. Input the non-calibrated values
and calibrated values of two measuring points (target 1 and target 2). Logic calibration
uses the following values.

T1-A: Target 1 display value before calibration
T1-B: Target 1 display value after calibration
T2-A: Target 2 display value before calibration
T2-B: Target 2 display value after calibration

Make settings so that the following conditions will be satisfied.
T1-A < T2-A and T1-B < T2-B
Furthermore, set the calibration span between 0.5 and 2.0.