Limit range – KEYENCE LS-7600 User Manual

Page 112

Advertising
background image

5-50

Chapter 5 Function Settings

5

Setting OUT1 Output

(1) Press the [ENT] key.

Setting OUT2 Output

(1) Press the [Down] key and move

the cursor frame to OUT2.

(2) Press the [ENT] key.

Limit Range

The following ranges are used for the limits.

Reference Value

HI output range: (Reference value + Upper limit) < Measured value
GO output range: (Reference value + Lower limit) <

= Measured value

<

= (Reference value

+ Upper limit)

LO output range: Measured value < (Reference value + Lower limit)

Possible Setting Conditions

A possible set value must be within a range where the following two items can be set.
Reference value + Upper limit value
Reference value + Lower limit value
In addition, the following condition must be satisfied.
Reference value + Upper limit value > Reference value + Lower limit value

Tips

Refer to Units, Default Minimum Values, and Setting Ranges on page 5-60 for
possible setting ranges in detail.

Threshold

HH output range: HH set value < Measured value
HI output range: HI set value < Measured value <

= HH set value

GO output range: LO set value <

= Measured value

<

= HI set value

LO output range: LL set value <

= Measured value < LO set value

LL output range: Measured value < LL set value

Possible Setting Conditions

HH set value > HI set value > LO set value > LL set value

Tips

Refer to Units, Default Minimum Values, and Setting Ranges on page 5-60 for
possible setting ranges in detail.

OUT1

UPPER

1.00000

STANDARD

10.00000

LOWER

—1.00000

OUT2

UPPER

1.00000

STANDARD

10.00000

LOWER

—1.00000

CAM1

PROG

P — 00

LIMIT

AREA
OUTPUT
CALIB

OPTION

A1 2
A2 2

01/02/18
17:47:30