Tolerance (LIMIT) Settings

The LS-7600 Series judges whether or not the measured values pass the criteria. The
range of values used for judgment is called a tolerance.Tolerances can be set to OUT1 and
OUT2 respectively.

Types of Limits

The following tolerance settings are possible.
• Reference value

Set the reference value and its upper and lower limits as deviation values. Measured
values can be compared to three levels (i.e., HI, GO, and LO). Use this mode if the
workpiece size changes but the deviations (tolerances) do not change.

• Threshold

Specify the upper and lower limits in absolute values. Measured values can be com-
pared to five levels (HH, HI, GO, LO, and LL).

Limit Setting Method

Take the following procedure if the remote control console is used.
Items appearing on the LCD monitor are for measuring head 1.

1

Set the [Slide Switch] key to PROG.

The LCD monitor changes to the PROGRAM SETUP screen from the RUN screen.

2

Press the [Down] key and move the
cursor to LIMIT.

3

Press the [ENT] key.

The cursor frame is displayed around
OUT1.

OUT1

UPPER

1.00000

STANDARD

10.00000

LOWER

—1.00000

OUT2

UPPER

1.00000

STANDARD

10.00000

LOWER

—1.00000

CAM1

PROG

P — 00

LIMIT

AREA
OUTPUT
CALIB

OPTION

A1 2
A2 2

01/02/18
17:47:30

OUT1

UPPER

1.00000

STANDARD

10.00000

LOWER

—1.00000

OUT2

UPPER

1.00000

STANDARD

10.00000

LOWER

—1.00000

CAM1

PROG

P — 00

LIMIT

AREA
OUTPUT
CALIB

OPTION

A1 2
A2 2

01/02/18
17:47:30

Advertising