Change of edge detection threshold – KEYENCE LS-7600 User Manual

Page 82

Advertising
background image

5-20

Chapter 5 Function Settings

5

Change of Edge Detection Threshold

The following section provides information on how to make threshold changes for the
measurement of highly transparent objects.

Transparent tape

Threshold
(50%)

50% threshold

Position

Position

Optical

reception

Brighter

Darker

Optical reception

Brighter

Darker

Light blocked
by tape

Measuring range

Detection impossible

Threshold
(80%)

80% threshold

Detection

possible

When measuring a highly transparent
target, such as a plastic tape, the
amount of light blocked by the tape is
minimal. Therefore, the tape may not be
detected if the threshold is low.

In that case, increase the threshold level
so that the tape will be detected

Setting Range

The threshold can be set between 10% and 99% in 1% increments.

Note:

• Only a single threshold can be set for each measuring head.
• In the measurement using the window edge such as T-EDGE or B-EDGE,

the measurement may become unstable with a higher threshold. In such a
case, use the reference edge or edge 0.

Optical Reception Waveform Output

While making a threshold setting for the target in the measuring area, an oscilloscope
can be used to output the optical reception waveform of the target so that the threshold
can be adjusted while monitoring the waveform.

Threshold level
(OUT2 analog output on terminal 23)

Optical reception waveform
(OUT1 analog output on terminal 22)

Approx. 450 ms

Approx.

10 ms

10V

Approx. 6.5 V

0V

Trigger pulse

Advertising