Focus Monitor
The focus monitor is a function that makes it possible to check whether the target is lo-
cated within the range where the highest precision measurement of the target is ensured.
The focus monitor range is the distance between the receiver and target in the measuring
area. When the target is within the range specified, the area setting on the LCD monitor will
display “F.”
FOCUS display is lit
The center of the measuring area
Focus display is OFF
Setting Range
The value is set at 10 levels between 0 and 9. Set the value to 0 if the focus monitor
function is not used. Refer to Focus Characteristics on page 9-6 for the set values and
focus range in detail.
AREA1
HEAD
1
AREA
DIA
A-DISP
ON
FOCUS 5
A-TEST OFF
Tips
The result of focus monitoring can be externally output as function output.
Refer to Function Output Selection on page 5-19 for details.
The focus monitor displays and outputs the average value for 256
measurements.
AREA1
HEAD
1
AREA
DIA
A-DISP
ON
FOCUS
5
A-TEST OFF
■
Setting Method
1
Select FOCUS and press the [ENT]
key.
The cursor moves to the setting item.
2
Press the [Up or Down] key and input
the set value.
3
Press the [ENT] key.
The input item is entered and the cursor
moves to FOCUS.