Chapter 3 Easy Setting Guide

Measurement of Outer Diameter and Width (with Two Measuring Heads for Larger Objects)

A sheet cannot be located in the measuring area of a single measuring head if the width or
diameter of the sheet is excessively large. In that case, the sheet can be measured with
two measuring heads as shown below with the automatic zero function and the master
workpiece used. The result of output is displayed as OUT1.

L

Set Values

Set item

Calibration

Tolerance setting

Area setting

Option setting

Output

setting

AREA

Application mode

Operation

Averaging

RUN mode

Offset

Set value

AREA 1

No settings are required.

Make settings if required.

A1: +1

A2: +1

Select the value according to the line speed.

Normal

Input the size of master workpiece.

No settings are required.

No settings are required.

Set the upper and lower limit values.

AREA 2

OUT 1

OUT 2

DUAL

Reference

Automatic Zero Function: Measure the master workpiece and make automatic zero
settings. (Page 4-10)
Option Settings (Page 5-55)
Area Settings (Page 5-9)
Calibration Settings (Page 5-22)
Output Settings (Page 5-29)
Tolerance Settings (Page 5-49)

L= (X+Y)

L=(X+Y)

When measuring the width of a
sheet

When measuring the outer
diameter of a round bar