6 bus states – Measurement Computing Analyzer488 User Manual

Page 107

Advertising
background image

Section 4

Serial Controller

4.13

mode allows control of the IEEE 488 bus using high level commands.
Low Level Controller mode allows direct control over all data and bus
management lines using an alternate command set.

4.11.6 Bus States

This section describes the bus command and data transfers using

IEEE 488 bus mnemonics abbreviated as follows:

DIO lines

8 7

6 5

4 3

2 1

ATN

Attention

data

Data String

DCL

Device Clear

x 0

0 1

0 1

0 0

EOI

End Or Identify

GET

Group Execute Trigger

x 0

0 0

1 0

0 0

GTL

Go To Local

x 0

0 0

0 0

0 1

IFC

Interface Clear

LAG

Listen Address Group

x 0

1 a

d d

r

n

LLO

Local Lock Out

x 0

0 1

0 0

0 1

MLA

My Listen Address

x 0

1 a

d d

r

n

MTA

My Talk Address

x 1

0 a

d d

r

n

PPC

Parallel Poll Configure

x 0

0 0

0 1

0 1

PPD

Parallel Poll Disable

x 1

1 1

0 0

0 0

PPE

Parallel Poll Enable

x 1

1 0

S P3 P2 P1

PPU

Parallel Poll Unconfigure

x 0

0 1

0 1

0 1

REN

Remote Enable

SDC

Selected Device Clear

x 0

0 0

0 1

0 0

SPD

Serial Poll Disable

x 0

0 1

1 0

0 1

SPE

Serial Poll Enable

x 0

0 1

1 0

0 0

SRQ

Service Request

TAG

Talker Address Group

x 1

0 a

d d

r

n

TCT

Take Control

x 0

0 0

1 0

0 1

UNL

Unlisten

x 0

1 1

1 1

1 1

UNT

Untalk

x 1

0 1

1 1

1 1

(x = "don't care")