Measurement Computing Analyzer488 User Manual
Page 185
Section 4
Serial Controller
4.91
valid data, a bus error, or the trigger point. The second two digits
show the state of the bus data lines.
REMARKS
For VIEW BUS, format is as follows:
Byte 1
Byte 2
(bit positions)
(bit positions)
(Control and Handshake Lines)
(Data Lines)
8 7 6 5
4 3
2 1
8 7 6 5 4 3 2 1
REN IFC SRQ EOI DAV NRFD NDAC ATN D8
D1
For VIEW location, format is as follows:
Byte 1
Byte 2
(bit positions)
(bit positions)
(Control and Status Lines)
(Data Lines)
8 7 6 5
4 3
2 1
8 7 6 5 4 3 2 1
REN IFC SRQ EOI BERR VALIDTRIG ATN
D8
D1
MODE Any
BUS STATES
None
QUERY
None
RESPONSE
EXAMPLES
PRINT#1,"CONTROLLER ON"
PRINT#1,"CLEAR 20"
PRINT#1,"VIEW BUS"
INPUT#1,A$;PRINT A$
Response: BUS 0304
PRINT#1,"ERASE"
PRINT#1,"CONTROLLER ON"
PRINT#1,"RECORD ON"
PRINT#1,"CLEAR 20"
PRINT#1,"RECORD OFF"
PRINT#1,"VIEW 2"
INPUT#1,A$;PRINT A$
Response: 00002 0534