Measurement Computing Analyzer488 User Manual
Page 151
Section 4
Serial Controller
4.57
count
is the number of record memory locations to be printed,
from 0 to 32768.
RESPONSE
Response from ? [location] is :
<location> <bus command or ASCII data> < data (hex)>
< data (decimal)> < data (binary) > <EOI state> <ATN state>
<SRQ state> <REN state> <IFC state> <ERROR (if bus error)>
Response from ? BUS is :
< BUS > <bus command or ASCII data> < data (hex)>
< data (decimal)> < data (binary) > <EOI state> <ATN state>
<SRQ state> <REN state> <IFC state> <DA ND NR (if all
handshake lines are asserted)>
MODE Any
BUS STATES
None
QUERY
None
RESPONSE
EXAMPLE
PRINT#1,"ERASE"
PRINT#1,"CONTROLLER ON"
PRINT#1,"RECORD ON"
PRINT#1,"CLEAR 20"
PRINT#1,"RECORD OFF"
PRINT#1,"?3"
Response is fixed field:
00003 SDC 04 004 0000 0100 ATN
PRINT#1,"?BUS"
returns:
BUS SDC 04 004 0000 0100 ATN ND