Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 102

Advertising
background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

84

6.7.2.7. Number of Records

The number of data records in the GFC 7000TA is limited to about a cumulative one million data points in
all channels (one megabyte of space on the disk-on-chip). However, the actual number of records is also
limited by the total number of parameters and channels and other settings in the DAS configuration.
Every additional data channel, parameter, number of samples setting etc. and will reduce the maximum
amount of data points somewhat. In general, however, the maximum data capacity is divided amongst all
channels (max: 20) and parameters (max: 50 per channel).

The DAS will check the amount of available data space and prevent the user from specifying too many
records at any given point. If, for example, the DAS memory space can accommodate 375 more data
records, the ENTR key will disappear when trying to specify more than that number of records. This
check for memory space may also make an upload of an DAS configuration with APICOM or a Terminal
program fail, if the combined number of records would be exceeded. In this case, it is suggested to either
try from the front panel what the maximum number of records can be or use trial-and-error in designing
the DAS script or calculate the number of records using the DAS or APICOM manuals. To set the
number of records for one channel from the front panel, press SETUP-DAS-EDIT-ENTR and the following
touchscreen button sequence.


Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 1 2, 900


PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC

<SET

SET>

EDIT

PRINT

EXIT

SETUP X.X

NUMBER OF RECORDS:000

<SET SET>

EDIT

PRINT

EXIT

SETUP X.X

REPORT PERIODD:DAYS:0

0 0 0 0 0 ENTR EXIT

ENTR

accepts the new

setting and returns to the

previous menu.

EXIT

ignores the new setting

and returns to the previous

menu.

Toggle buttons to set

number of records

(1-99999)

SETUP X.X

EDIT RECOPRDS (DELET DATA)

YES NO

NO returns to the

previous menu.

YES will delete all data

in this channel.

Press SET> key until…

From the DATA ACQUISITION menu

(see Section 6.12.2.2)

Advertising