Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 138

Advertising
background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

120

6.13. SETUP

MORE DIAG: Using the Diagnostics

Functions

A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu. As these
parameters are dependent on firmware revision (see Menu Tree A-6 in Appendix A). The individual
parameters, however, are explained in more detail in the indicated in 6-15. These tools can be used in a
variety of troubleshooting and diagnostic procedures and are referred to in many places of the
maintenance and trouble-shooting s.

Table 6-15: GFC 7000TA Diagnostic (DIAG) Functions

DIAGNOSTIC FUNCTION AND MEANING

Front Panel

Mode Indicator

SIGNAL I/O

: Allows observation of all digital and analog signals in

the instrument. Allows certain digital signals such as valves and
heaters to be toggled ON and OFF.

DIAG I/O

6.13.2

ANALOG OUTPUT

: When entered, the analyzer performs an analog

output step test. This can be used to calibrate a chart recorder or
to test the analog output accuracy.

DIAG AOUT

6.13.3

ANALOG I/O CONFIGURATION

: the signal levels of the instruments

analog outputs may be calibrated (either individually or as a
group). Various electronic parameters such as signal span, and
offset are available for viewing and configuration.

DIAG AIO

6.13.4

ELECTRIC TES

T: The analyzer is performing an electric test. This

test simulates IR detector signal in a known manner so that the
proper functioning of the sync/demod board can be verified.

DIAG OPTIC

6.13.5

DARK CALIBRATION

: The analyzer is performing a dark

calibration procedure. This procedure measures and stores the
inherent dc offset of the sync/demod board electronics.

DIAG ELEC

6.13.6

PRESSURE CALIBRATION

: The analyzer records the current

output of the sample gas pressure sensor. This value is used by
the CPU to compensate the CO

2

concentration.

DIAG PCAL

6.13.7

FLOW CALIBRATION

: This function is used to calibrate the gas flow

output signal. This setting is retained when exiting DIAG.

DIAG FCAL

6.13.8

TEST CHAN OUTPUT

: Configures the A4 analog output channel.

DIAG TCHN

6.13.9

Advertising