ESD

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

261

11.4.2.5. Packing Components for Return to Teledyne Instruments Customer

Service.

Always pack electronic components and assemblies to be sent to Teledyne Instruments Customer
Service in anti-ESD bins, tubes or bags.

WARNING

DO NOT

use pink-poly bags.

NEVER

allow any standard plastic packaging materials to touch the electronic

component/assembly directly

This includes

,

but is not limited to, plastic bubble-pack, Styrofoam peanuts,

open cell foam, closed cell foam, and adhesive tape

DO NOT use standard adhesive tape as a sealer. Use ONLY anti-ESD tape

1. Never carry the component or assembly without placing it in an anti-ESD bag or bin.

2. Before using the bag or container allow any surface charges on it to dissipate:

 If you are at the instrument rack, hold the bag in one hand while your wrist strap is connected to

a ground point.

 If you are at an anti-ESD workbench, lay the container down on the conductive work surface.
 In either case wait several seconds.

3. Place the item in the container.

4. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.

 Folding the open end over isolates the component(s) inside from the effects of static fields.
 Leaving the bag open or simply stapling it shut without folding it closed prevents the bag from

forming a complete protective envelope around the device.

NOTE

If you do not already have an adequate supply of anti-ESD bags or containers available, Teledyne

Instruments’ Customer Service department will supply them. Follow the instructions listed above

for working at the instrument rack and workstation.


: