C bus, Touchscreen interface, Lcd display module – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 260

Advertising
background image

Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

242

A voltmeter should be used to verify that the DC voltages are correct per the values in the table below,
and an oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate if the supplies are
producing excessive noise (> 100 mV p-p).

Table 10-7: DC Power Supply Acceptable Levels

POWER

SUPPLY

ASSY

VOLTAG

E

CHECK RELAY BOARD TEST POINTS

MIN V

MAX V

FROM TEST POINT

TO TEST POINT

NAME # NAME #

PS1 +5 Dgnd

1 +5 2 4.8

5.25

PS1 +15 Agnd 3 +15 4 13.5 16V
PS1 -15 Agnd 3 -15V 5 -14V -16V
PS1 Agnd Agnd 3 Dgnd 1 -0.05 0.05
PS1 Chassis Dgnd 1 Chassis N/A -0.05 0.05
PS2 +12

+12V

Ret

6 +12V 7 11.75 12.5

PS2 Dgnd

+12V

Ret

6 Dgnd 1 -0.05 0.05

10.5.3. I

2

C Bus

Operation of the I

2

C bus can be verified by observing the behavior of D1 on the Relay PCA & D2 on the

Valve Driver PCA . Assuming that the DC power supplies are operating properly, the I

2

C bus is operating

properly if: D1 on the relay PCA and D2 of the Valve Driver PCA are flashing

There is a problem with the I

2

C bus if both D1 on the relay PCA and D2 of the Valve Driver PCA are

ON/OFF constantly (steady lit or steady off)

10.5.4. Touchscreen Interface

Verify the functioning of the touchscreen by observing the display when pressing a touchscreen control
button. Assuming that there are no wiring problems and that the DC power supplies are operating
properly, if pressing a control button on the display does not change the display, any of the following may
be the problem:

 The touchscreen controller may be malfunctioning.
 The internal USB bus may be malfunctioning.

You can verify this failure by logging on to the instrument using APICOM or a terminal program to any of
the communications ports. If the analyzer responds to remote commands and the display changes
accordingly, the touchscreen interface may be faulty.

10.5.5. LCD Display Module

Verify the functioning of the front panel display by observing it when power is applied to the instrument.
Assuming that there are no wiring problems and that the DC power supplies are operating properly, the
display screen should light and show the splash screen with logo and other indications of its state as the
CPU goes through its initialization process.