Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

23

3.5.4. Connecting the Status Outputs

If you wish utilize the analyzer’s status outputs to interface with a device that accepts logic-level digital
inputs, such as programmable logic controllers (PLC’s) they are accessed via a 12-pin connector on the
analyzer’s rear panel labeled STATUS.

STATUS

1 2 3 4 5 6 7 8 D

+

NOTE

Most PLC’s have internal provisions for limiting the current the input will draw.

When connecting to a unit that does not have this feature, external resistors must

be used to limit the current through the individual transistor outputs to ≤50mA

(120 Ω for 5V supply).

The pin assignments for the status outputs can be found in the table below:

Table 3-5: Status Output Pin-outs

OUTPUT #

STATUS

DEFINITION

CONDITION

1

SYSTEM OK

On if no faults are present.

2

CONC VALID

On if CO

2

concentration measurement is valid.

If the CO

2

concentration measurement is invalid, this bit is OFF.

3

HIGH RANGE

On if unit is in high range of DUAL or AUTO range modes.

4

ZERO CAL

On whenever the instruments ZERO point is being calibrated.

5

SPAN CAL

On whenever the instruments SPAN point is being calibrated.

6

DIAG MODE

On whenever the instrument is in DIAGNOSTIC mode.

7

ALARM1

On whenever the measured CO

2

concentration is above the set point for

ALM1

8

ALARM2

On whenever the measured CO

2

concentration is above the set point for

ALM2

D

EMITTER BUS

The emitters of the transistors on pins 1-8 are bused together.

+

DC POWER

+ 5 VDC

Digital Ground

The ground level from the analyzer’s internal DC power supplies.