Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 172

Advertising
background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

154

6.15.4.2. Hessen com Port Configuration

Hessen protocol requires the communication parameters of the GFC 7000TA’s com ports to be set
differently than the standard configuration as shown in the table below.

Table 6-29: RS-232 Communication Parameters for Hessen Protocol

Parameter Standard Hessen

Data Bits

8

7

Stop Bits

1

2

Parity None

Even

Duplex Full

Half

To change the rest of the com port parameters. See Section 6.11.5.

To change the baud rate of the GFC 7000TA’s com ports, See Section 6.11.8.

NOTE

Make sure that the communication parameters of the host computer are also

properly set.

Also, the instrument software has a 200 ms. Latency before it responds to

commands issued by the host computer. This latency should present no problems,

but you should be aware of it and not issue commands to the instrument too

frequently.