5.4.2. Ambient Zero/Ambient Span Valve

This valve option is intended for applications where zero air is supplied by a zero air generator like a
bottle of N

2

and span gas are being supplied by Gas Dilution Calibrator like the Teledyne Instruments

Model 700 or 702. Internal zero/span and sample/cal valves control the flow of gas through the
instrument, but because the calibrator limits the flow of span gas no shutoff valve is required.

In order to ensure that span gas does not migrate backwards through the vent line and alter the
concentration of the span gas, a gas line not less than 2 meters in length should be attached to the vent
span outlet on the rear panel of the analyzer. To prevent the buildup of back pressure, this vent line
should not be greater than 10 meters in length.

The following table describes the state of each valve during the analyzer’s various operational modes.

Table 5-2: Ambient Zero/Ambient Span Valve Operating States

Mode Valve

Condition

SAMPLE

(Normal State)

Sample/Cal

Open to SAMPLE inlet

Zero/Span

Open to IZS inlet

ZERO CAL

Sample/Cal

Open to ZERO/SPAN valve

Zero/Span

Open to IZS inlet

SPAN CAL

Sample/Cal

Open to ZERO/SPAN valve

Zero/Span

Open to PRESSURE SPAN inlet