Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

97

6.9. SETUP

PASS: Password Feature

The GFC 7000TA provides password protection of the calibration and setup functions to prevent
unauthorized adjustments. When the passwords have been enabled in the PASS menu item, the system
will prompt the user for a password anytime a password-protected function is requested.

There are three levels of password protection, which correspond to operator, maintenance, and
configuration functions. Each level allows access to all of the functions in the previous level.

Table 6-9: Password Levels

PASSWORD

LEVEL

MENU ACCESS ALLOWED

No password

Operator

TEST, MSG, CLR

818

Maintenance

CAL, CALZ, CALS

929

Configuration

SETUP, VARS, DIAG


To enable or disable passwords, press the following keystroke sequence: