Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 104
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
86
6.7.2.11. Disabling/Enabling Data Channels
Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-chip.
The ALL_01 channel of the GFC 7000TA, for example, is disabled by default.
To disable a data channel, follow the instruction shown in Section 6.7.2.2 then press:
Edit Data Channel Menu
SETUP X.X
0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
CHANNEL ENABLE:ON
<SET SET>
EDIT
EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
CHANNEL ENABLE:ON
OFF
ENTR EXIT
Toggle to turn
channel ON or OFF
Press SET> key until…
From the DATA ACQUISITION menu
(see Section 6.7.2.2)