Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

72

6.7.1. DAS Structure

The DAS is designed around the feature of a “record”. A record is a single data point of one parameter,
stored in one (or more) data channels and generated by one of several triggering event. The entire DAS
configuration is stored in a script, which can be edited from the front panel or downloaded, edited and
uploaded to the instrument in form of a string of plain-text lines through the communication ports.
DAS data are defined by the PARAMETER type and are stored through different triggering EVENTS in
data CHANNELS, which relate triggering events to data parameters and define certain operational
functions related to the recording and reporting of the data.

6.7.1.1. DAS Channels

The key to the flexibility of the DAS is its ability to store a large number of combinations of triggering
events and data parameters in the form of data channels. Users may create up to 20 data channels and
each channel can contain one or more parameters. For each channel one triggering event is selected
and up to 50 data parameters, which can be the same or different between channels. Each data channel
has several properties that define the structure of the channel and allow the user to make operational
decisions regarding the channel (Table 6-7).

Table 6-7: DAS Data Channel Properties

PROPERTY DESCRIPTION DEFAULT

SETTING

RANGE

NAME

The name of the data channel.

“NONE”

Up to 6 letters and digits

(more with APICOM, but

only the first six are

displayed on the front

panel).

TRIGGERING

EVENT

The event that triggers the data channel to measure
and store its data parameters. See APPENDIX A-5
for a list of available triggering events.

ATIMER

See Appendix A-5 For a

complete list.

NUMBER &

PARAMETER

LIST

A user-configurable list of data types to be recorded
in any given channel. See APPENDIX A-5 for a list
of available parameters

1 – DETMES

See Appendix A-5 For a

complete list.

STARTING

DATE

The starting date when a channel starts collecting
data

01-JAN-03

Any actual date in the past

or future.

SAMPLE

PERIOD

The amount of time between each data point that is
averaged into one mean reported every REPORT
PERIOD.

000:01:00

000:00:01 to 366:23:59

(Days:Hours:Minutes)

REPORT

PERIOD

The amount of time between each channel data
point.

000:01:00

000:00:01 to

366:23:59

(Days:Hours:Minutes)

NUMBER OF

RECORDS

The number of reports that will be stored in the data
file. Once the specified limit has been exceeded,
the oldest data are over-written to make space for
new data.

100

1 to 1 million, limited by

available storage space.

RS-232

REPORT

Enables the analyzer to automatically report channel
values to the RS-232 ports.

OFF

OFF or ON

CHANNEL
ENABLED

Enables or disables the channel. Provides a
convenient means to temporarily disable a data
channel.

ON

OFF or ON

CAL HOLD OFF

Disables sampling of data parameters while
instrument is in calibration mode.
When enabled here – there is also a length of the
DAS HOLD OFF after calibration mode, which is set
in the VARS menu.

OFF

OFF or ON

(Section 6.7.2.11.)