Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

169

STEP TWO:

Set the expected CO

2

Span Gas concentration:

This sequence causes the

analyzer to prompt for the

expected CO

2

span

concentration.

M-P CAL

CO2 SPAN CONC: 450.000 Conc

0 0 0

4 5 .0

ENTR

EXIT

M-P CAL RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > ZERO

CONC

EXIT

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST >

CAL

CALZ CALS

SETUP

The CO

2

span

concentration values

automatically default to

400.0 Conc

.

To change this value to

the actual concentration of

the span gas, enter the

number by pressing the

button under each digit

until the expected value

appears.

EXIT

ignores the new setting

and returns to the previous

display.

ENTR

accepts the new setting

and returns to the

previous display.

NOTE

For this Initial Calibration it is important to independently verify the PRECISE CO

2

Concentration Value of the SPAN gas.

If the source of the Span Gas is from a Calibrated Bottle, use the exact

concentration value printed on the bottle.

STEP THREE:

Perform the zero/span calibration. Zero and span checks using the zero/span valve

option are similar to that described in Section 7.2, except that zero air and span gas is supplied to the
analyzer through the zero/span valves rather than through the sample inlet port.

The zero and cal operations are initiated directly and independently with dedicated buttons (CALZ &
CALS

).