Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 240

Advertising
background image

Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

222

Table 10-2: Test Functions - Indicated Failures

TEST

FUNCTIONS

INDICATED FAILURE(S)

TIME

Time of day clock is too fast or slow: To adjust See Section 6.6.
Battery in clock chip on CPU board may be dead.

RANGE

Incorrectly configured measurement range(s) could cause response problems with a Data logger or chart
recorder attached to one of the analog output.
If the Range selected is too small, the recording device will over range.
If the Range is too big, the device will show minimal or no apparent change in readings.

STABIL

Indicates noise level of instrument or CO

2

concentration of sample gas (See Section 10.4.2 for causes).

CO2 MEAS

&

CO2 REF

If the value displayed is too high the IR Source has become brighter. Adjust the variable gain
potentiometer on the sync/demod board
If the value displayed is too low or constantly changing and the CO

2

REF

is OK:

o

Failed multiplexer on the mother board

o

Failed sync/demod board

o

Loose connector or wiring on sync/demod board

Flow of purge gas to the GFC wheel housing may have stopped

If the value displayed is too low or constantly changing and the CO

2

REF

is BAD:

o

GFC wheel stopped or rotation is too slow

o

Failed sync/demod board IR source

o

Failed IR source

o

Failed relay board

o

Failed

I

2

C

bus

o

Failed IR photo-detector

MR RATIO

When the analyzer is sampling zero air and the ratio is too low:

o

The reference cell of the GFC wheel is contaminated or leaking.

o

The alignment between the GFC wheel and the segment sensor, the M/R sensor or both is

incorrect.

o

Failed sync/demod board

o

Flow of purge gas to the GFC wheel housing may have stopped

When the analyzer is sampling zero air and the ratio is too high:

o

Zero air is contaminated

o

Failed IR photo-detector

PRES

See Table 11-1 for SAMPLE PRES WARN

SAMPLE FL

Check for gas flow problems. see Section 11.1.6

SAMPLE

TEMP

SAMPLE TEMP

should be close to BENCH TEMP. Temperatures outside of the specified range or

oscillating temperatures are cause for concern

BENCH TEMP

Bench temp control improves instrument noise, stability and drift. Temperatures outside of the specified
range or oscillating temperatures are cause for concern. See Table 11-1 for BENCH TEMP WARNING

WHEEL TEMP

Wheel temp control improves instrument noise, stability and drift. Outside of set point or oscillating
temperatures are cause for concern. See Table 11-1 for WHEEL TEMP WARNING

BOX TEMP

If the box temperature is out of range, check fan in the power supply module. Areas to the side and rear
of instrument should allow adequate ventilation. See Table 11-1 for BOX TEMP WARNING.

PHT DRIVE

If this drive voltage is out of range it may indicate one of several problems:
- A poor mechanical connection between the various components in inside the detector housing
- An electronic failure of the IR Photo-Detector’s built-in cooling circuitry, or;
- A temperature problem inside the analyzer chassis. In this case other temperature warnings would also
be active such as BENCH TEMP WARNING or BOX TEMP WARNING.

Advertising