Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

145

6.15.1.2. Control Inputs

These inputs allow the user to remotely initiate Zero and Span calibrations. Two methods for energizing
the inputs is provided below; the first using the internal +5V available on the CONTROL IN connector and
the second, if an external, isolated supply is employed.

Table 6-25: Control Input Pin Assignments

INPUT STATUS

CONDITION

WHEN

ENABLED

A

EXTERNAL ZERO

CAL

Zero calibration mode is activated. The mode field of the display will
read ZERO CAL R.

B

EXTERNAL SPAN

CAL

Span calibration mode is activated. The mode field of the display will
read SPAN CAL R.

C

Unused

D

Unused

E

Unused

F

Unused

DIGITAL GROUND

Provided to ground an external device (e.g., recorder).

U

DC power for Input

pull ups

Input for +5 VDC required to activate inputs A - F. This voltage can be
taken from an external source or from the “+” pin.

+

Internal +5V Supply

Internal source of +5V which can be used to activate inputs when
connected to pin U.

There are two methods to activate control inputs. The internal +5V available from the “+” pin is the most
convenient method (Figure 6-18). However, to ensure that these inputs are truly isolated, a separate,
external 5 VDC power supply should be used.

CONTROL IN

A B C D E F U +

S
P

A
N

Z
E

R

O

CONTROL IN

A B C D E F U +

-

+

5 VDC Power

Supply

S
P
A
N

Z
E

R
O

Local Power Connections

External Power Connections

Figure 6-14:

Control Inputs