Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 46

Advertising
background image

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

28

7. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more

than 2 meters long.

NOTE

Ideally, the pressure of the sample gas should be at ambient pressure (0 psig).

Maximum pressure of sample gas should not exceed 1.5 in-Hg over ambient.

In applications where the sample gas is received from a pressurized manifold, a vent

must be placed as shown to equalize the sample gas with ambient atmospheric

pressure before it enters the analyzer.

This vent line must be:

At least 0.2m long
No more than 2m long and vented outside the shelter or immediate area

surrounding the instrument.

8. Attach sources of zero air and span gas (see Figures 3-3 through 3-8 inclusive).

 Span Gas is a gas specifically mixed to match the chemical composition of the type of gas

being measured at near full scale of the desired measurement range.

In the case of CO

2

measurements made with the Teledyne Instruments Model GFC 7000TA

Analyzer it is recommended that you use a gas calibrated to have a CO

2

content equaling

80% of the range of compositions being measured.

EXAMPLE: If the application is to measure between 0 ppm and 500 ppm, an appropriate Span
Gas would be 400 ppm. If the application is to measure between 0 ppm and 100 ppm, an
appropriate Span Gas would be 80 ppm.

 Span Gas can be purchased in pressurized canisters or created using Dynamic Dilution

Calibrator such as the Teledyne Instruments Model T700 and a source of dried air scrubbed
of CO

2

such as a bottle of N

2

.

 Zero Air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all

components that might affect the analyzer’s readings.

In the case of CO

2

measurements this means less than 0.1 ppm of CO

2

and Water Vapor

(when dew point <-15˚C). Zero Air can be purchased in pressurized canisters.

9. Attach an exhaust line to the exhaust outlet port.

 The exhaust from the pump and vent lines should be vented to atmospheric pressure using

maximum of 10 meters of ¼” PTEF tubing.

CAUTION

Venting should be outside the shelter or immediate area surrounding the instrument.

Advertising