Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 109
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
91
To select the Analog Output Range Type press:
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL
DUAL AUTO
ENTR EXIT
EXIT
Returns
to the Main
SAMPLE
Display
Go To
Section
6.7.3
Go To
Section
6.7.4
Go To
Section
6.7.5
Only one of the
range modes may
be active at any
time.
NOTE
Upper span limit setting for the individual range modes are shared. Resetting the
span limit in one mode also resets the span limit for the corresponding range in
the other modes as follows:
SNGL
DUAL
AUTO
Range
Range1 Low
Range
Range2
High