Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

122

To enter the signal I/O test mode, press:

SAMPLE RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

EXAMPLE

SAMPLE

ENTER SETUP PASS: 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

PREV NEXT

JUMP

ENTR

EXIT

DIAG I / O

Test Signals Displayed Here

PREV NEXT

JUMP

PRNT EXIT

Use JUMP to go

directly to a specific

signal location

See Appendix A-4 for

a complete list of

available SIGNALS

DIAG I / O

JUMP TO: 12

1 2

ENTR

EXIT

EXAMPLE:

Enter 12 to Jump to

12)

ST_CONC_VALID

DIAG I / O

ST_CONC_VALID = ON

PREV NEXT

JUMP

ON

PRNT

EXIT

Exit to return

to the

DIAG

menu

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

Use NEXT & PREV to

move between signal

types.

Pressing PRNT will send a formatted printout to the serial port and can be captured

with a computer or other output device.

EXIT

returns

to the main

SAMPLE

display

6.13.3. Analog Output Step Test

This test can be used to check the accuracy and proper operation of the analog outputs. The test forces
all four analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20%
increments. This test is useful to verify the operation of the data logging/recording devices attached to
the analyzer.