Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 56
Getting Started
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
38
2. Set the expected CO
2
span gas concentration
This sequence causes the
analyzer to prompt for the
expected CO
2
span
concentration.
M-P CAL
CO2 SPAN CONC: 400.000 Conc
0 0 0
4 5 .0
ENTR
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > ZERO
CONC
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST >
CAL
SETUP
The CO
2
span
concentration values
automatically default to
400.0 Conc
.
To change this value to
the actual concentration of
the span gas, enter the
number by pressing the
button under each digit
until the expected value
appears.
EXIT
ignores the new setting
and returns to the previous
display.
ENTR
accepts the new setting
and returns to the
previous display..
NOTE
For this Initial Calibration it is important to independently verify the precise CO
2
Concentration Value of the SPAN gas.
If the source of the Span Gas is from a Calibrated Bottle, use the exact
concentration value printed on the bottle.