Step 2 – activate o, Concentration to the o, Sensor output, press – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 59
Getting Started
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
41
SAMPLE
GAS TO CAL:CO2
CO2
O2
ENTR EXIT
SAMPLE RANGE = 500.000 PPM CO2 =XXX.X
< TST TST > CAL
SETUP
M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX
<TST TST> ZERO SPAN
CONC
EXIT
SAMPLE
GAS TO CAL:O2
NOX
O2
ENTR
EXIT
M-P CAL O2 SPAN CONC:20.8%
0
2
0
.8
0
ENTR EXIT
The O
2
span concentration value automatically defaults to
20.8 %
.
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the O
2
calibration gases.
EXIT
ignores the new
setting and returns to
the previous display.
ENTR
accepts the new
setting and returns to
the previous menu.
STEP 2 – ACTIVATE O
2
SENSOR STABILITY FUNCTION
To change the stability test function from NO
x
concentration to the O
2
sensor output, press: