Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual
Page 199
Advertising
Maintenance
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
181
Table 8-1: GFC 7000TA Maintenance Schedule
ITEM ACTION FREQ
CAL
CHECK
REQ’D.
MANUAL
DATE PERFORMED
Particulate
Filter
Replace
Weekly or as
needed
Yes
Verify Test
Functions
Record and
analyze
Weekly or after
any
Maintenance
or Repair
No
Pump
Diaphragm
Replace Annually Yes
Perform
Flow Check
Check Flow
Annually
No
Perform
Leak Check
Verify Leak
Tight
Annually or
after any
Maintenance
or Repair
No
Pneumatic
lines
Examine
and clean
As needed
Yes if
cleaned
Cleaning Clean As
needed
Only if
cover
remv’d
Advertising