Specifications and approvals, Specifications – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 23

Advertising
background image

Specifications

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

5

2. SPECIFICATIONS AND APPROVALS

2.1. Specifications

Table 2-1: Model GFC 7000TA Basic Unit Specifications

GFC 7000TA Parameter

GFC 7000TA Specification

Ranges
(Physical Analog Output)

Min: 0-2 ppm Full scale
Max: 0-2,000 ppm Full scale
Selectable, dual ranges and auto ranging supported

Measurement Units

ppb, ppm, µg/m

3

, mg/m

3

, % (user selectable)

Zero Noise

< 0.1 ppm (RMS)

Span Noise

< 1% of reading (RMS)

Lower Detectable Limit

< 0.2 ppm

1

Zero Drift (24 hours)

<0.25 ppm

1

Span Drift (24 hours)

<0.5% of reading

1

Lag Time

10 seconds

Rise/Fall Time

<60 seconds to 95%

Linearity

1% of full scale

Precision

0.5% of reading

Sample Flow Rate

800cm

3

/min. ±10%

O

2

Sensor option adds 110 cm³/min, ±20%, to total flow through when installed.

Temperature Coefficient

< 0.1% of Full Scale per

o

C

Voltage Coefficient

< 0.05% of Full Scale per V

AC Power Rating

100V-120V, 220V – 240 V, 50/60 Hz

Analog Output Ranges

10V, 5V, 1V, 0.1V (selectable)

Analog Output Resolution

1 part in 4096 of selected full-scale voltage

Recorder Offset

±10%

Standard I/O

1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs
6 opto-isolated digital control inputs (3 defined, 3 spare)
4 analog outputs

Optional I/O

1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs

Environmental Installation

category (over-voltage category) II; Pollution degree 2

Temperature Range

5-40

o

C

Humidity Range

0 - 95% RH, non-condensing

Dimensions H x W x D

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Advertising