Electric test, Meas and co – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 154

Advertising
background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

136

6.13.5. Electric Test

The electric test function substitutes simulated signals for CO

2

MEAS

and CO

2

REF

, generated by

circuitry on the sync/demod board, for the output of the IR photo-detector. While in this mode the user
can also view the same test functions viewable from the main SAMPLE display. When the test is
running, the concentration reported on the front panel display should be 40.0 ppm.

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

Press <TST TST> to view Test Functions

NOTE: CO

2

MEAS

and CO

2

REF

will be artificially altered

to enforce a CO

2

reading of 40.0 ppm.

All other Test Functions will report the correct operational

value

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Exit returns

to the

DIAG

Menu

Repeat Pressing NEXT unti . . .

DIAG

ELECTRIC TEST

PREV NEXT

ENTR

EXIT

DIAG ELEC RANGE=50.000 PPM

CO2= 40.0


<TST TST>

EXIT

SAMPLE RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR

EXIT

Advertising