6.13.5. Electric Test
The electric test function substitutes simulated signals for CO
2
MEAS
and CO
2
REF
, generated by
circuitry on the sync/demod board, for the output of the IR photo-detector. While in this mode the user
can also view the same test functions viewable from the main SAMPLE display. When the test is
running, the concentration reported on the front panel display should be 40.0 ppm.
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM
EXIT
Press <TST TST> to view Test Functions
NOTE: CO
2
MEAS
and CO
2
REF
will be artificially altered
to enforce a CO
2
reading of 40.0 ppm.
All other Test Functions will report the correct operational
value
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Exit returns
to the
DIAG
Menu
Repeat Pressing NEXT unti . . .
DIAG
ELECTRIC TEST
PREV NEXT
ENTR
EXIT
DIAG ELEC RANGE=50.000 PPM
CO2= 40.0
<TST TST>
EXIT
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT