Zero/span calibration on auto range or dual ranges – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 190

Advertising
background image

Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

172

7.5.1. Zero/Span Calibration on Auto Range or Dual Ranges

If the analyzer is being operated in dual range mode or auto range mode, then the high and low ranges
must be independently calibrated.

When the analyzer is in either dual or auto range modes the user must run a separate calibration
procedure for each range. After pressing the CAL, CALZ or CALS buttons the user is prompted for the
range that is to be calibrated as seen in the CALZ example below:

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST

TST

>

CAL

CALZ

CALS SETUP

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

Wait until STABIL

falls below 1.0 ppm.

This may take several

minutes.

SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X

<TST

TST>

CAL

CALZ CALS SETUP

SAMPLE

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST >

CAL

CALZ CALS SETUP

Set the Display to show the

STABIL

test function.

This function calculates the

stability of the CO

2

measurement

Continue Calibration as per

Standard Procedure

ZERO CAL M

RANGE = 500.000 PPM CO2 =XXX.X

< TST TST >

ZERO

SPAN

CONC

EXIT

ANALYZER ENTERS

ZERO CAL MODE

SAMPLE

RANGE TO CAL: LOW

LOW

HIGH

ENTR SETUP

SAMPLE

RANGE TO CAL: HIGH

LOW HIGH

ENTR

SETUP

Once this selection is made, the calibration procedure continues as previously described in Section 7.2.
The other range may be calibrated by starting over from the main SAMPLE display.

Advertising