Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

95

6.8.6. Range Units

The GFC 7000TA can display concentrations in parts per billion (10

9

mols per mol, PPB), parts per million

(10

6

mols per mol, PPM), micrograms per cubic meter (µg/m

3

, UG), milligrams per cubic meter (mg/m

3

,

MG

) or percent (volume CO

2

/volume sample gas, %). Changing units affects all of the display, analog

outputs, COM port and DAS values for all reporting ranges regardless of the analyzer’s range mode.

NOTE

Concentrations displayed in mg/m3 and ug/m3 use 0C, 760 mmHg for Standard

Temperature and Pressure (STP). Consult your local regulations for the STP used

by your agency.

Conversion factors from volumetric to mass units are:

CO

2

: ppb x 1.96 = µg/m3; ppm x 1.96 = mg/m3

To change the concentration units:

SETUP X.X

CONC UNITS: %

PPM PPB

UGM

MGM %

ENTER EXIT

SAMPLE RANGE = 500.00 PPB CO2=X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP X.X

CONC UNITS: PPM

PPM PPB

UGM

MGM

%

ENTER

EXIT

Select the preferred

concentration unit.

EXIT

returns

to the main menu.

ENTR

accepts

the new unit,

EXIT

returns

to the SETUP

menu.

NOTE

Once the units of measurement have been changed the unit MUST be recalibrated,

as the “expected span values” previously in effect will no longer be valid. Simply

entering new expected span values without running the entire calibration routine

is not sufficient.