Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

158

6.15.4.7. Setting Hessen Protocol Status Flags

Teledyne Instruments’ implementation of Hessen protocols includes a set of status bits that the
instrument includes in responses to inform the host computer of its condition. Each bit can be assigned
to one operational and warning message flag. The default settings for these bit/flags are:

Table 6-31: Default Hessen Status Bit Assignments

STATUS FLAG NAME

DEFAULT BIT

ASSIGNMENT

WARNING FLAGS

SAMPLE FLOW WARNING

0001

BENCH TEMP WARNING

0002

SOURCE WARNING

0004

BOX TEMP WARNING

0008

WHEEL TEMP WARNING

0010

SAMPLE TEMP WARNING

0020

SAMPLE PRESSURE WARNING

0040

INVALID CONC
(The Instrument’s Front Panel Display Will Show The
Concentration As “XXXX”)

0080

OPERATIONAL FLAGS

Instrument Off

0100

In Manual Calibration Mode

0200

In Zero Calibration Mode

0400

In Span Calibration Mode

0800

UNITS OF MEASURE FLAGS

UGM 0000

MGM 2000

PPB 4000

PPM 6000

SPARE/UNUSED BITS

100, 1000, 8000

UNASSIGNED FLAGS (0000)

Sync Warning

Relay Board Warning

Conc Alarm 1

Analog Cal Warning

Conc Alarm 2

Cannot Dyn Zero

Photo Temp Warning

Cannot Dyn Span

System Reset

Invalid Conc

Rear Board Not Detected

NOTES:

It is possible to assign more than one flag to the same Hessen status bit. This

allows the grouping of similar flags, such as all temperature warnings, under the
same status bit.
Be careful not to assign conflicting flags to the same bit as each status bit will be

triggered if any of the assigned flags is active.