Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

98

Example: If all passwords are enabled, the following keypad sequence would be required to enter the
SETUP menu:

SAMPLE RANGE = 500.000 PPM CO2 =X.XXX


< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS: 0


0 0 0

ENTR EXIT

prompts for password

number

See Table 6-8 for

Passwords and Levels

Example: this

password enables the

SETUP mode

SAMPLE

ENTER SETUP P ASS: 0


8 1 8

ENTR EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

Press individual

buttons to set

numbers

Note that the instrument still prompts for a password when entering the VARS and DIAG menus, even if
passwords are disabled, but it displays the default password (818) upon entering these menus. The user
only has to press ENTR to access the password-protected menus but does not have to enter the required
number code.