Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

175

To program the Sequence:

Toggle

buttons to

set

number of

days

between

procedures

(1-367)

Toggle

buttons to set
delay time for

each iteration

of the

sequence:

HH:MM

(0 – 24:00)

SETUP X.X

DELTA DAYS: 1


<SET SET>

EDIT

EXIT

SETUP X.X

DELTA DAYS: 1


0 0 2

ENTR

EXIT

SETUP X.X

DELTA DAYS:2


<SET

SET>

EDIT

EXIT

SETUP X.X

DELTA TIME00:00


<SET SET>

EDIT

EXIT

SETUP X.X

DELTA TIME: 00:00


0 0 :3 0

ENTR

EXIT

SETUP X.X

DELTA TIEM:00:30


<SET

SET>

EDIT

EXIT

SETUP X.X

STARTING TIME:14:15


<SET

SET>

EDIT

EXIT

EXIT

returns

to the SETUP

Menu

SETUP X.X

DURATION:15.0 MINUTES


<SET SET>

EDIT

EXIT

SETUP X.X

DURATION 15.0MINUTES


3 0 .0

ENTR

EXIT

SETUP X.X

DURATION:30.0 MINUTES


<SET

SET>

EDIT

EXIT

Toggle

buttons to

set

duration for

each

iteration of

the

sequence:

Set in

Decimal

minutes

from

0.1 – 60.0

SETUP X.X

CALIBRATE: OFF


<SET SET>

EDIT

EXIT

SETUP X.X C ALIBRATE: OFF

ON

ENTR

EXIT

SETUP X.X

CALIBRATE: ON


<SET SET> EDIT

EXIT

Toggle

button

between

Off and

ON

SETUP X.X

SEQ 2) ZERO–SPAN, 2:00:30


PREV NEXT

MODE

SET

EXIT

Sequence

Delta Time

MODE Delta Days

SETUP X.X

PRIMARY SETUP MENU


CFG

ACAL

DAS RNGE PASS CLK MORE EXIT

SETUP X.X

SEQ 1) DISABLED


NEXT

MODE

EXIT

SETUP X.X

SEQ 2) DISABLED


PREV NEXT

MODE

EXIT

SETUP X.X

MODE: DISABLED


NEXT

ENTR EXIT

SETUP X.X

MODE: ZERO


PREV

NEXT

ENTR EXIT

SETUP X.X

MODE: ZERO–SPAN


PREV

NEXT

ENTR

EXIT

SETUP X.X

SEQ 2) ZERO–SPAN, 1:00:00


PREV NEXT

MODE

SET

EXIT

SETUP X.X

TIMER ENABLE: ON


SET>

EDIT

EXIT

SETUP X.X

STARTING DATE: 01–JAN–02


<SET SET>

EDIT

EXIT

SETUP X.X

STARTING DATE: 01–JAN–02


0

4 SEP

0

3

ENTR

EXIT

Toggle

buttons to set

day, month &

year:

Format :

DD-MON-Y

Y

SETUP X.X

STARTING DATE: 04–SEP–03


<SET

SET>

EDIT

EXIT

Default

value is

ON

SETUP X.X

STARTING DATE: 04–SEP–03


<SET

SET>

EDIT

EXIT

SETUP X.X

STARTING TIME:00:00


<SET SET>

EDIT

EXIT

SETUP X.X

STARTING TIME:00:00


1 4 : 1 5

ENTR

EXIT/

Toggle buttons

to set time:

Format : HH:MM

This is a 24 hr

clock .

PM hours are

13 – 24.

Example

2:15 PM = 14:15

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL CALZ CALS

SETUP