Initial calibration procedure – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 55

Advertising
background image

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

37

3.8. Initial Calibration Procedure

The next task is to calibrate the analyzer.

To perform the following calibration you must have sources for zero air and span gas available for input
into the sample port on the back of the analyzer. See Section 3.1.3 for instructions for connecting these
gas sources.

While it is possible to perform this procedure with any range setting we recommend that you perform this
initial checkout using the 500 ppm range.

NOTE

The following procedure assumes that the instrument does not have any of the available

Zero/Span Valve Options installed.

See Section 7.4 for instructions for calibrating instruments possessing Z/S valve options.

1. Set the Analog Output Range:

SETUP X.X

CFG DAS

RNGE

PASS CLK MORE

EXIT

SAMPLE RANGE = 500.000 PPM

CO2

=X.XXX

< TST TST > CAL

SETUP

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

Press this button to set

the analyzer for

SNGL

DUAL

or AUTO ranges

SETUP X.X

RANGE: 500.000 CONC

0 0

5

0 0 .

0

ENTR EXIT

To change the value of the

reporting range span, enter the

number by pressing the key under
each digit until the expected value

appears.

Press this button to select the

concentration units of measure:

PPB, PPM, UGM, MGM

EXIT

ignores the new setting and

returns to the RANGE CONTROL

MENU

.

ENTR

accepts the new setting and

returns to the

RANGE CONTROL MENU

.

SETUP X.X

RANGE:

500.000 Conc

0 0 0 5 0 .0

ENTR

EXIT