Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

26

3.6.1. Basic Pneumatic Connections

NOTE

In order to prevent dust from getting into the gas flow channels of your analyzer,

it was shipped with small plugs inserted into each of the pneumatic fittings on the

back panel. Remove these plugs before operating the analyzer. It is recommended

that they be stored for future use (moving, storing or shipping the analyzer).

Figure 3-4 illustrates the basic configuration for gas supply and exhaust lines to the Model GFC 7000TA
Analyzer. Figure 3-5 illustrates the internal gas flow of the instrument in its basic configuration.

Please refer to Figure 3-2 for pneumatic connections at the rear panel and Table 3-2 for description.

NOTE

Sample and calibration gases should only come into contact with PTFE (Teflon),

FEP, glass, stainless steel or brass.

Table 3-7: Rear Panel Pneumatic Connections

REAR PANEL LABEL

FUNCTION

SAMPLE

Connect a gas line from the source of sample gas here.
Calibration gasses are also inlet here on units without zero/span valve or IZS

options installed.

EXHAUST

Connect an exhaust gas line of not more than 10 meters long here.

PRESSURE SPAN

On units with zero/span valve options installed, connect a gas line to the source

of calibrated span gas here.

VENT SPAN

Span gas vent outlet for units with zero/span valve options installed.
Connect an exhaust gas line of not more than 10 meters long here.

IZS

Internal zero air scrubber.

On units with zero/span valve options installed but no internal zero air
scrubber, attach a gas line to the source of zero air here.

PURGE IN

This inlet supplies purge air to the GFC wheel housing (see Section 10.2.3)
Connect a source of dried air that has been scrubbed of CO