Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

140

6.13.9. Test Channel Output

When activated, output channel A4 can be used to report one of the test functions viewable from the
SAMPLE mode display. To activate the A4 channel and select a test function, follow this button
sequence:

SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR

EXIT

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

DIAG TCHN

TEST CHANNEL: NONE

NEXT

ENTR

EXIT

DIAG

TEST CHAN OUTPUT

PREV

NEXT

ENTR

EXIT

Continue to press NEXT until …

DIAG TCHN

TEST CHANNEL: CO2 MEASURE

PREV

NEXT

ENTR

EXIT

Press PREV or NEXT

to move through the

list of available

parameters

(Table 6-13)

Press EXIT to

return to the

DIAG

menu

Press ENTR to select

the displayed

parameter activating

the test channel.

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

Table 6-22: Test Parameters Available for Analog Output A4

TEST CHANNEL

ZERO

FULL SCALE

NONE

Test Channel is turned off

CO

2

MEASURE

0 mV

5000 mV*

CO

2

REFERENCE

0 mV

5000 mV*

SAMPLE PRESS

0 "Hg

40 "Hg

SAMPLE FLOW

0 cc/m

1000 cc/m

SAMPLE TEMP

0

C 70C

BENCH TEMP

0

C 70C

WHEEL TEMP

0

C 70C

CHASSIS TEMP

0

C 70C

PHT DRIVE

0 mV

5000 mV