Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

126

6.13.4.1. Analog Output Signal Type and Range Span Selection

To select an output signal type (DC Voltage or current) and level for one output channel, activate the
ANALOG I/O CONFIGURATION MENU (See Section 6.13.1) then press:

DIAG AIO

CONC_OUT_2 RANGE: 5V

SET>

EDIT

EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED: NO

< SET

SET>

CAL

EXIT

FROM ANALOG I/O CONFIGURATION MENU

DIAG AIO OUTPUT RANGE:

5V

0.1V 1V 5V

10V

CURR

ENTR

EXIT

To set the signal

level and type of

the selected

channel

DIAG AIO

CONC_OUT_2:5V, CAL

< SET SET>

EDIT

EXIT

Press SET> to select the
analog output channel to be

configured. Press EDIT to

continue

Pressing ENTR records the new setting

and returns to the previous menu.

Pressing EXIT ignores the new setting and

returns to the previous menu.

DIAG AIO OUTPUT RANGE:

10V

0.1V 1V 5V

10V

CURR

ENTR EXIT

6.13.4.2. Analog Output Calibration Mode

The analog outputs can be calibrated automatically or manually. In its default mode, the instrument is
configured for automatic calibration of all channels. Manual calibration should be used for the 0.1V range
or in cases where the outputs must be closely matched to the characteristics of the recording device.
Outputs configured for automatic calibration can be calibrated as a group or individually. Calibration of
the analog outputs needs to be carried out on first startup of the analyzer (performed in the factory as part
of the configuration process) or whenever re-calibration is required.