Remote das configuration – Teledyne GFC-7000T - Trace CO2 Analyzer - manual User Manual

Page 106

Advertising
background image

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

Teledyne Analytical Instruments

88

6.7.3. Remote DAS Configuration

Editing channels, parameters and triggering events as described in this can be performed via the
APICOM remote control program using the graphic interface shown in Figure 6-5. Refer to Section 6.15
for details on remote access to the GFC 7000TA analyzer.

Figure 6-5:

APICOM User Interface for DAS Configuration

Once a DAS configuration is edited (which can be done offline and without interrupting DAS data
collection), it is conveniently uploaded to the instrument and can be stored on a computer for later review,
alteration or documentation and archival. Refer to the APICOM manual for details on these procedures.
The APICOM user manual (Teledyne Instruments part number 039450000) is included in the APICOM
installation file, which can be downloaded at http://www.teledyne-api.com/software/apicom/. It is
recommended that you start by downloading the default DAS configuration, getting familiar with its
command structure and syntax conventions, and then altering a copy of the original file offline before
uploading the new configuration.

CAUTION

Whereas the editing, adding and deleting of DAS channels and parameters of one

channel through the front-panel touchscreen can be done without affecting the

other channels, uploading an DAS configuration script to the analyzer through its

communication ports will erase all data, parameters and channels by replacing

them with the new DAS configuration. Backup of data and the original DAS

configuration is advised before attempting any DAS changes.

Advertising