3. GETTING STARTED

3.1. Unpacking and Initial Set Up

CAUTION

To avoid personal injury, always use two persons to lift and carry the Model GFC

7000TA.

WARNING

Never disconnect PCAs, wiring harnesses or electronic subassemblies while the

instrument is under power.

NOTE

It is recommended that you store shipping containers/materials for future use if/when the instrument should
be returned to the factory for repair and/or calibration service. See Warranty section in this manual and
shipping procedures on our

Website at:

http://www.teledyne-api.com

under Customer Support > Return Authorization.

1. Verify that there is no apparent external shipping damage. If damage has occurred, please

advise the shipper first, then Teledyne Instruments.

2. Included with your analyzer is a printed record of the final performance characterization

performed on your instrument at the factory. This record, titled Final Test and Validation Data
Sheet

(P/N 04596) is an important quality assurance and calibration record for this instrument. It

should be placed in the quality records file for this instrument.

3. Carefully remove the top cover of the analyzer and check for internal shipping damage.

 Remove the set-screw located in the top, center of the Front panel.
 Remove the 2 screws fastening the top cover to the unit (one per side towards the rear).
 Slide the cover backwards until it clears the analyzer’s front bezel.
Lift the cover straight up.



CAUTION – Avoid Warranty Invalidation

Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too
small to be felt by the human nervous system. Damage resulting from failure to
use ESD protection when working with electronic assemblies will void the
instrument warranty.

See A Primer on Electro-Static Discharge in this manual for more information on
preventing ESD damage.

Advertising